Sagittal Skin Contour of the Cervical Spine: Inter-examiner and Intra-examiner Reliability of the Flexicurve Instrument
Sagittal Skin Contour of the Cervical Spine: Inter-examiner and Intra-examiner Reliability of the Flexicurve Instrument
Authors
- Deed E. Harrison, DC
- Jason W. Haas, DC
- Donald D. Harrison, PhD, DC
- Burt Holland, PhD
- Tadeusz Janik, PhD
Publication
Journal of Manipulative and Physiological Therapeutics 2005; 28(7):516-9
Article Link
Abstract
Objectives: To evaluate reliability of a simple instrument, the flexicurve, in determining cervical sagittal skin contour.
Methods: This study obtained repeated random measurements involving 3 investigators and 30 subjects once per day over a 2-day trial period. Thirty normal subjects were examined for cervical spine skin contour twice by 3 separate investigators with a 1-day delay. With subjects in a neutral standing position, investigators placed the flexicurve on the
posterior portion of the subject’s neck from the external occipital protuberance to the vertebral prominens and traced the flexicurve shape onto paper. The tracings were divided into 6 equal arcs and digitized. Statistical computation was performed on the depth at 5 points, arc angle, and arc radius of curvature. Interexaminer and intraexaminer correlation
coefficients (ICCs) were calculated to determine reliability.
Results: All interexaminer correlation coefficients were in the poor range (b0.40). For the arc radius, arc angle, depth at top one third, and depth at bottom two thirds, the intraexaminer correlation coefficients were in the poor range. For the 3 deepest depths, the intraexaminer correlation coefficients were in the fair range (0.4-0.50).
Conclusion: The flexicurve showed marginal reliability with most (12/16) ICCs in the poor range (ICC b 0.40) and 4 values in the fair range (0.4 b ICC b 0.5). (J Manipulative Physiol Ther 2005;28:516- 519)